Molecular imaging of small laboratory animals, such as mice and rats, is a vital tool in the study of disease. Mouse models of human diseases, such as cancer, Parkinson's disease, and Alzheimer's disease, are providing important clues to the causes, diagnosis and treatment of these, and many other, disorders. Ultra-high resolution PET and SPECT have enabled the visualization of biochemical processes in vivo in small animals. Despite the important results obtained using pinhole SPECT, it is clear that the sensitivity of SPECT systems is sub-optimal. Some sensitivity can be recovered by either increasing the pinhole diameter, resulting in a loss of resolution, or increasing the injected dose, which may lead to radiation damage in the subject. This project will develop novel, widely applicable, non-invasive methods for small animal SPECT, using multiple-pinhole techniques to improve the image statistics, while maintaining the spatial resolution, and potentially reducing the radiation dose to the animal. Various multiple-pinhole configurations will be investigated, including both overlapping (in which the projection images overlap on the detector face), and non-overlapping (in which the geometric arrangement of pinholes is designed to produce separate images on the detector). Monte Carlo computer simulations will be used to optimize the pinhole arrangement, providing high-resolution images, without compromising the signal-to-noise ratio or introducing image artifacts. The optimum design will be application-dependent, and determined using quantitative techniques and Hotelling observer studies. Additional image reconstruction improvements, such as correction for depth-of-interaction in the crystal, and pinhole septal edge penetration, will be investigated using Monte Carlo simulations. Model-based iterative reconstruction techniques will be developed to deconstruct the data from the multiple-pinhole systems, and to incorporate the corrections for pinhole edge penetration and depth of- interaction. The optimized multiple-pinhole system will be constructed and compared against the single pinhole device in both phantom and animal studies, to validate the multiple-pinhole approach, and to test the reconstruction algorithm. The improvements in the imaging technology proposed in this application will lead to enhanced sensitivity, and improvements in the throughput of small animal molecular imaging due to the decreased imaging time required. It also takes advantage of adapting existing clinical SPECT machines, rather than developing a dedicated small animal imaging system, which makes it much less expensive and more widely available. ? ?

Agency
National Institute of Health (NIH)
Institute
National Institute of Biomedical Imaging and Bioengineering (NIBIB)
Type
Research Project (R01)
Project #
1R01EB001809-01
Application #
6708705
Study Section
Special Emphasis Panel (ZRG1-EB (52))
Program Officer
Pastel, Mary
Project Start
2003-09-10
Project End
2007-07-31
Budget Start
2003-09-10
Budget End
2004-07-31
Support Year
1
Fiscal Year
2003
Total Cost
$356,625
Indirect Cost
Name
University of Pennsylvania
Department
Radiation-Diagnostic/Oncology
Type
Schools of Medicine
DUNS #
042250712
City
Philadelphia
State
PA
Country
United States
Zip Code
19104
Thakur, Mathew L; Devadhas, Devakumar; Zhang, Kaijun et al. (2010) Imaging spontaneous MMTVneu transgenic murine mammary tumors: targeting metabolic activity versus genetic products. J Nucl Med 51:106-11
Mukherjee, Archana; Wickstrom, Eric; Thakur, Mathew L (2009) Imaging oncogene expression. Eur J Radiol 70:265-73
Thakur, Mathew L (2009) Genomic biomarkers for molecular imaging: predicting the future. Semin Nucl Med 39:236-46
Petruzzi, Nicholas; Shanthly, Nylla; Thakur, Mathew (2009) Recent trends in soft-tissue infection imaging. Semin Nucl Med 39:115-23
Thomas, Daniel; Bal, Harshali; Arkles, Jeffrey et al. (2008) Noninvasive assessment of myocardial viability in a small animal model: comparison of MRI, SPECT, and PET. Magn Reson Med 59:252-9
Shen, Dinggang; Liu, Dengfeng; Cao, Zixiong et al. (2007) Coregistration of magnetic resonance and single photon emission computed tomography images for noninvasive localization of stem cells grafted in the infarcted rat myocardium. Mol Imaging Biol 9:24-31
Acton, Paul D; Thomas, Daniel; Zhou, Rong (2006) Quantitative imaging of myocardial infarct in rats with high resolution pinhole SPECT. Int J Cardiovasc Imaging 22:429-34
Accorsi, R; Metzler, S D (2006) Non-diverging analytic expression for the on-axis sensitivity of converging collimators: experimental verification. Phys Med Biol 51:5697-705
Bal, Girish; Acton, Paul D (2006) Analytical derivation of the point spread function for pinhole collimators. Phys Med Biol 51:4923-50
Accorsi, R; Metzler, S D (2006) Non-diverging analytic expression for the on-axis sensitivity of converging collimators: analytic derivation. Phys Med Biol 51:5675-96

Showing the most recent 10 out of 14 publications