Currently, medical CT scanners are under rapid development with an increasingly larger cone angle, while biomedical micro-CT scanners are already in cone-beam geometry. Despite the importance of cone-beam CT, cone-beam image reconstruction algorithms are not fully developed. There is a critical and immediate need for a dynamic volumetric performance of cone-beam CT, subject to multiple constraints such as dose, noise, range, contrast, etc. The overall goal of this project is to develop and optimize analytic cone-beam algorithms with an emphasis on high temporal resolution and short scan range, and directly applicable to major applications such as cardiac imaging, CT fluoroscopy, perfusion studies, CT angiography, oncologic imaging, small animal imaging, as well as PET and SPECT. This project is based on the latest cone-beam CT results, and focuses on both approximate and exact reconstruction in the Feldkamp-type, Grangeat-type and Katsevich-type frameworks respectively.
The specific aims are to (1) improve Feldkamp-type algorithms for less than half-scan data by scanning pattern design and weighting scheme optimization; (2) extend Grangeat-type half-scan algorithms for long object reconstruction by correcting cone-beam data, and transform the Radon space based reconstruction into the filtered backprojection format; (3) modify Katsevich-type algorithms for dynamic reconstruction by detection coverage minimization and n-PI geometry-based formulation; and (4) evaluate and validate the proposed cone-beam algorithms in theoretical analysis, numerical simulation and phantom experiments, and demonstrate their feasibility and utilities in mouse and patient studies. On completion, superior and practical cone-beam algorithms will have been systematically developed with excellent image quality for dynamic volumetric CT and micro-CT. These proposed algorithms will have been implemented on a PC cluster. The advantages of the algorithms will have been demonstrated in mouse and patient studies.

Agency
National Institute of Health (NIH)
Institute
National Institute of Biomedical Imaging and Bioengineering (NIBIB)
Type
Research Project (R01)
Project #
5R01EB002667-04
Application #
7100181
Study Section
Special Emphasis Panel (ZRG1-SRB (51))
Program Officer
Mclaughlin, Alan Charles
Project Start
2003-09-30
Project End
2006-11-09
Budget Start
2006-08-01
Budget End
2006-11-09
Support Year
4
Fiscal Year
2006
Total Cost
$105,017
Indirect Cost
Name
University of Iowa
Department
Radiation-Diagnostic/Oncology
Type
Schools of Medicine
DUNS #
062761671
City
Iowa City
State
IA
Country
United States
Zip Code
52242
Lu, Yang; Katsevich, Alexander; Zhao, Jun et al. (2010) Fast exact/quasi-exact FBP algorithms for triple-source helical cone-beam CT. IEEE Trans Med Imaging 29:756-70
Yu, Hengyong; Wang, Ge (2010) A soft-threshold filtering approach for reconstruction from a limited number of projections. Phys Med Biol 55:3905-16
Wang, Ge; Yu, Hengyong (2010) Can interior tomography outperform lambda tomography? Proc Natl Acad Sci U S A 107:E92-3, author reply E94-5
Lu, Yang; Yu, Hengyong; Cao, Guohua et al. (2010) Multibeam field emission x-ray system with half-scan reconstruction algorithm. Med Phys 37:3773-81
Yang, Jiansheng; Yu, Hengyong; Jiang, Ming et al. (2010) High Order Total Variation Minimization for Interior Tomography. Inverse Probl 26:350131-3501329
Wang, Ge; Yu, Hengyong; Ye, Yangbo (2009) A scheme for multisource interior tomography. Med Phys 36:3575-81
Bharkhada, Deepak; Yu, Hengyong; Ge, Shuping et al. (2009) Cardiac computed tomography radiation dose reduction using interior reconstruction algorithm with the aorta and vertebra as known information. J Comput Assist Tomogr 33:338-47
Lu, Yang; Zhao, Jun; Wang, Ge (2009) Exact image reconstruction with triple-source saddle-curve cone-beam scanning. Phys Med Biol 54:2971-91
Ye, Lena; Yu, Hengyong; Wang, Ge (2009) Determination of exact reconstruction regions in composite-circling cone-beam tomography. Med Phys 36:3448-54
Bharkhada, Deepak; Yu, Hengyong; Dixon, Robert et al. (2009) Demonstration of dose and scatter reductions for interior computed tomography. J Comput Assist Tomogr 33:967-72

Showing the most recent 10 out of 51 publications