There is a growing interest in the use of chronic deep brain stimulation (DBS) of the thalamus, globus pallidus, and the subthalamic nucleus for the treatment of medically refractory movement disorders and other neurological and psychiatric conditions. The number of patients undergoing DBS surgeries is anticipated to increase rapidly. Use of magnetic resonance imaging (MRI) is crucial for the management of patients with DBS. Critical studies include verification of lead position, assessment of patients with poor outcomes, evaluation for other intracranial pathologies, and performing functional MRI studies. However, electronically activated devices are generally considered to be contraindicated for patients undergoing MRI, based on reports of adverse effects in the literature. Our primary hypothesis is that parameters can be defined that will enable safe MR imaging in patients with DBS. Building upon our preliminary data, We intend to conduct a comprehensive safety analysis, functional assessment of the stimulation devices, and evaluation of the artifacts produced by the DBS. We will use 1.5 Tesla and 3Tesla MR systems, body radio frequency (RF) and head RF coils to conduct all experiments. Specifically, in-vitro experiments will be performed to assess DBS magnetic field interactions (translational attraction and torque), MRI-related heating (including computer modeling), induced electrical currents, and identification of any effects on device functionality. The artifacts produced by the DBS electrodes will also be quantified. The identification of safe MRI parameters will enable patients with neurostimulation systems to routinely undergo diagnostic MR procedures and allow the use of this important imaging modality for the optimal localization of DBS electrodes.

Agency
National Institute of Health (NIH)
Institute
National Institute of Neurological Disorders and Stroke (NINDS)
Type
Research Project (R01)
Project #
5R01NS044575-03
Application #
6795463
Study Section
Special Emphasis Panel (ZNS1-SRB-A (02))
Program Officer
Pancrazio, Joseph J
Project Start
2002-08-15
Project End
2006-07-31
Budget Start
2004-08-01
Budget End
2006-07-31
Support Year
3
Fiscal Year
2004
Total Cost
$469,559
Indirect Cost
Name
Cleveland Clinic Lerner
Department
Surgery
Type
Schools of Medicine
DUNS #
135781701
City
Cleveland
State
OH
Country
United States
Zip Code
44195
Park, Sung-Min; Kamondetdacha, Rungkiet; Nyenhuis, John A (2007) Calculation of MRI-induced heating of an implanted medical lead wire with an electric field transfer function. J Magn Reson Imaging 26:1278-85
Baker, Kenneth B; Tkach, Jean A; Phillips, Micheal D et al. (2006) Variability in RF-induced heating of a deep brain stimulation implant across MR systems. J Magn Reson Imaging 24:1236-42
Baker, Kenneth B; Nyenhuis, John A; Hrdlicka, Greg et al. (2005) Neurostimulation systems: assessment of magnetic field interactions associated with 1.5- and 3-Tesla MR systems. J Magn Reson Imaging 21:72-7
Bhidayasiri, Roongroj; Bronstein, Jeff M; Sinha, Shantanu et al. (2005) Bilateral neurostimulation systems used for deep brain stimulation: in vitro study of MRI-related heating at 1.5 T and implications for clinical imaging of the brain. Magn Reson Imaging 23:549-55
Baker, Kenneth B; Tkach, Jean; Hall, John D et al. (2005) Reduction of magnetic resonance imaging-related heating in deep brain stimulation leads using a lead management device. Neurosurgery 57:392-7; discussion 392-7
Rezai, Ali R; Phillips, Michael; Baker, Kenneth B et al. (2004) Neurostimulation system used for deep brain stimulation (DBS): MR safety issues and implications of failing to follow safety recommendations. Invest Radiol 39:300-3
Baker, Kenneth B; Tkach, Jean A; Nyenhuis, John A et al. (2004) Evaluation of specific absorption rate as a dosimeter of MRI-related implant heating. J Magn Reson Imaging 20:315-20
Sharan, A; Rezai, A R; Nyenhuis, J A et al. (2003) MR safety in patients with implanted deep brain stimulation systems (DBS). Acta Neurochir Suppl 87:141-5