Computed tomography (CT) provides important medical benefits, but for patient safety it is essential that CT providers use the lowest dose of radiation consistent with achieving the needed diagnostic performance. New algorithmic approaches to image reconstruction will be critical to reducing the dose without compromising image quality; however, the development of novel approaches to image reconstruction is hampered because many image scientists do not have access to CT projection data from patient exams. We propose to develop data sets, metrics, and software tools that will help investigators create and compare new approaches to dose reduction and will guide clinical users in selecting optimized scanning parameters and reconstruction methods.
In Aim 1, we will make reference patient data sets available to researchers in a standardized format after removal of proprietary information. These data will include projection data, statistical noise maps, reconstructed images, and clinical reference information (validation of diagnosis and location, abstracted patient history) for common CT exams, as well as data simulating lower exposure levels. These data sets will greatly expand the pool of researchers that can develop and evaluate algorithms, and will permit comparison of competing approaches. The gold standard for measuring diagnostic performance, observer performance studies, is however very expensive and time consuming.
In Aim 2, we will develop highly automated, interactive, and freely available software tools that will facilitate rapid completion of observer performance studies in order to efficiently and meaningfully compare alternative scanning protocols and reconstruction methods. Still, because of the rapid pace of technical innovation, a substitute for efficient observer performance studies is essential to rapidly translate advances in dose reduction into patient care. Although task-based image quality metrics using model observers are attractive for this purpose, they have not been demonstrated to correlate with radiologist performance in clinical CT imaging.
In Aim 3, we will determine model observers that are substantially correlated with human observer performance in patient data for three common diagnostic tasks and for linear and non-linear image reconstruction techniques. Finally, quantitative methods are needed to assist clinical practices in choosing scanning protocol parameters that will achieve the required level of diagnostic performance using the lowest radiation dose.
In Aim 4, we will develop tools to efficiently and quantitatively optimize CT scanning protocols for specific diagnostic tasks. These tools will calculate quantitative measures of task-based image quality from easily performed phantom scans and will recommend scanning protocol parameters that will deliver the closest match to the desired level of diagnostic performance using the lowest radiation dose. This research is highly innovative and significant in that it will provide the CT community with novel data, methods, and software tools for objective evaluation and efficient optimization of scanning protocol parameters and emerging dose reduction approaches.

Public Health Relevance

While computed tomography (CT) provides important medical benefits, it is essential that CT exams use the lowest dose of radiation that allows accurate diagnosis. In this work, we will develop data sets, methods, and software tools that will help investigators achieve a significant reduction in CT dose without unknowingly compromising diagnostic accuracy.

Agency
National Institute of Health (NIH)
Institute
National Institute of Biomedical Imaging and Bioengineering (NIBIB)
Type
Research Project--Cooperative Agreements (U01)
Project #
4U01EB017185-04
Application #
9134142
Study Section
Special Emphasis Panel (ZEB1-OSR-D (M1))
Program Officer
Shabestari, Behrouz
Project Start
2013-09-01
Project End
2017-08-31
Budget Start
2016-09-01
Budget End
2017-08-31
Support Year
4
Fiscal Year
2016
Total Cost
$902,011
Indirect Cost
$324,790
Name
Mayo Clinic, Rochester
Department
Type
DUNS #
006471700
City
Rochester
State
MN
Country
United States
Zip Code
55905
Fletcher, Joel G; Fidler, Jeff L; Venkatesh, Sudhakar K et al. (2018) Observer Performance with Varying Radiation Dose and Reconstruction Methods for Detection of Hepatic Metastases. Radiology 289:455-464
Gong, Hao; Yu, Lifeng; Leng, Shuai et al. (2018) Correlation between model observers in uniform background and human observers in patient liver background for a low-contrast detection task in CT. Proc SPIE Int Soc Opt Eng 10577:
Hanson, G Jay; Michalak, Gregory J; Childs, Robert et al. (2018) Low kV versus dual-energy virtual monoenergetic CT imaging for proven liver lesions: what are the advantages and trade-offs in conspicuity and image quality? A pilot study. Abdom Radiol (NY) 43:1404-1412
Ferrero, Andrea; Chen, Baiyu; Li, Zhoubo et al. (2017) Technical Note: Insertion of digital lesions in the projection domain for dual-source, dual-energy CT. Med Phys 44:1655-1660
McCollough, Cynthia H; Bartley, Adam C; Carter, Rickey E et al. (2017) Low-dose CT for the detection and classification of metastatic liver lesions: Results of the 2016 Low Dose CT Grand Challenge. Med Phys 44:e339-e352
Favazza, Christopher P; Ferrero, Andrea; Yu, Lifeng et al. (2017) Use of a channelized Hotelling observer to assess CT image quality and optimize dose reduction for iteratively reconstructed images. J Med Imaging (Bellingham) 4:031213
Ferrero, Andrea; Favazza, Christopher P; Yu, Lifeng et al. (2017) Practical implementation of Channelized Hotelling Observers: Effect of ROI size. Proc SPIE Int Soc Opt Eng 10132:
Yu, Lifeng; Chen, Baiyu; Kofler, James M et al. (2017) Correlation between a 2D channelized Hotelling observer and human observers in a low-contrast detection task with multislice reading in CT. Med Phys 44:3990-3999
Zeng, Gengsheng L (2017) A fast method to emulate an iterative POCS image reconstruction algorithm. Med Phys 44:e353-e359
Chen, Yang; Liu, Jin; Xie, Lizhe et al. (2017) Discriminative Prior - Prior Image Constrained Compressed Sensing Reconstruction for Low-Dose CT Imaging. Sci Rep 7:13868

Showing the most recent 10 out of 28 publications