As recognized in the FOA for sub-mSv CT, the increased CT utilization and the associated increase in population radiation dose motivates the development of dose reduction methods. Any meaningful strategies for dose reduction must be coupled with an analysis of image quality. However, the relationship between dose and image quality is complex due to dependencies on the imaging task and patient-specific characteristics. The pro- posed effort develops task-driven CT through hardware modifications that permit customizing the CT acquisition to the patient, and through a task-based performance prediction framework that is used to drive optimal dose utilization for specific imaging scenarios. Specifically, a novel lightweight and compact x-ray beam modulator capable of high dynamic range modulations and suitable for typical CT gantry rotation rates will be developed based on multiple aperture devices (MADs). This approach allows for a high degree of control over the spatial profile of the x-ray beam including flattening of fluence profile arriving at the detector nd region-of- interest (ROI) scanning. Actuation of these devices will be driven by an image quality plan based on a 3D scout volume and a task-based detectability framework that includes sophisticated models of the measurement physics, imaging task definitions, the human visual system, and the particular reconstruction approaches applied to the data. While dynamic beam modulation alone yields significant dose reductions, these advantages will be synergized with additional reductions through the use of advanced statistical reconstruction algorithms customized for beam modulated acquisitions. We hypothesize that a task-driven diagnostic CT scanner tailored to the specific imaging needs of the patient will provide large enough dose reductions that many body CT scanning scenarios can be driven to sub-mSv levels as targeted by this program announcement. The following Specific Aims are proposed to develop and investigate task-driven CT: 1) Develop dynamic beam modulation hardware for integration into diagnostic CT scanners. Design, characterize, and integrate the MAD modulators into CT acquisition systems. 2) Create a reconstruction framework for dynamically modulated CT acquisitions. Adapt both traditional and statistical reconstruction algorithms to beam modulated data. 3) Develop a performance prediction framework for dynamically modulated CT. A sophisticated physical model and task-based mathematical observer will be developed that predicts the shift-variant, patient-specific, and acquisition-dependent image quality. 4) Develop strategies for driving patient- and task-based beam modulations. Using image quality plans (including possibly shift-variant specification, e.g. ROI imaging) and the prediction framework prospectively design optimal beam modulations within dose constraints. 5) Assess patient- and task-specific beam-modulated CT. Evaluation outcome measures will include quantitative imaging performance metrics, absorbed dose measurements and dose maps based on Monte Carlo estimation, and an observer preference test using cadaver studies including relations to minimum dose protocols.

Public Health Relevance

Despite the great diagnostic utility of x-ray CT, increased use of CT over the past decade and incomplete attention to the need for x-ray technique modification based on the particular patient and imaging scenario has led to concerns about the radiation dose to individuals and to the patient population as a whole. We propose a novel task-based acquisition and reconstruction framework that leverages both new hardware allowing for customization of the spatial distribution of the x-ray beam placing exposure where it is needed and advanced model-based reconstruction software that maximizes image quality for low dose CT. These synergistic technologies are combined to open up a new paradigm in CT imaging whereby protocols are driven by an initial ultra-low dose 3D scout scan and a mathematical framework for prediction of imaging task performance so that CT studies can be tailored to the specific patient and imaging needs.

Agency
National Institute of Health (NIH)
Institute
National Institute of Biomedical Imaging and Bioengineering (NIBIB)
Type
Research Project--Cooperative Agreements (U01)
Project #
5U01EB018758-03
Application #
9066149
Study Section
Special Emphasis Panel (ZEB1)
Program Officer
Shabestari, Behrouz
Project Start
2014-09-15
Project End
2018-05-31
Budget Start
2016-06-01
Budget End
2017-05-31
Support Year
3
Fiscal Year
2016
Total Cost
Indirect Cost
Name
Johns Hopkins University
Department
Biomedical Engineering
Type
Schools of Medicine
DUNS #
001910777
City
Baltimore
State
MD
Country
United States
Zip Code
21205
Wang, Wenying; Gang, Grace J; Mao, Andrew et al. (2018) Volume-of-interest CT imaging with dynamic beam filtering using multiple aperture devices. Conf Proc Int Conf Image Form Xray Comput Tomogr 2018:213-217
Wang, W; Gang, G J; Siewerdsen, J H et al. (2018) Spatial Resolution and Noise Prediction in Flat-Panel Cone-Beam CT Penalized-likelihood Reconstruction. Proc SPIE Int Soc Opt Eng 10573:
Webster Stayman, J; Tilley 2nd, Steven (2018) Model-based Multi-material Decomposition using Spatial-Spectral CT Filters. Conf Proc Int Conf Image Form Xray Comput Tomogr 2018:102-105
Mao, Andrew; Shyr, William; Gang, Grace J et al. (2018) Dynamic beam filtering for miscentered patients. Proc SPIE Int Soc Opt Eng 10573:
Gang, Grace J; Mao, Andrew; Siewerdsen, Jeffrey H et al. (2018) Implementation and Assessment of Dynamic Fluence Field Modulation with Multiple Aperture Devices. Conf Proc Int Conf Image Form Xray Comput Tomogr 2018:47-51
Gang, Grace J; Stayman, J Webster (2018) Joint Optimization of Fluence Field Modulation and Regularization for Multi-Task Objectives. Proc SPIE Int Soc Opt Eng 10573:
Mathews, A J; Gang, G; Levinson, R et al. (2017) Experimental evaluation of dual Multiple Aperture Devices for Fluence Field Modulated X-Ray Computed Tomography. Proc SPIE Int Soc Opt Eng 10132:
Gang, Grace J; Siewerdsen, Jeffrey H; Webster Stayman, J (2017) Task-driven optimization of CT tube current modulation and regularization in model-based iterative reconstruction. Phys Med Biol 62:4777-4797
Gang, G J; Siewerdsen, J H; Stayman, J W (2017) Joint Optimization of Fluence Field Modulation and Regularization in Task-Driven Computed Tomography. Proc SPIE Int Soc Opt Eng 10132:
Gomes, Juliana; Gang, Grace J; Mathews, Aswin et al. (2017) An Investigation of Low-Dose 3D Scout Scans for Computed Tomography. Proc SPIE Int Soc Opt Eng 10132:

Showing the most recent 10 out of 17 publications