Despite the fact that SWL has been in clinical use for over fifteen years, the physical mechanisms for stone communication are still not well understood. However, there is now a growing body of experimental evidence and an evolving general acceptance by the clinical community, that SWL leads to some degree of permanent damage to treated kidneys. The mechanisms responsible for this tissue damage are also not well understood. We hypothesize that acoustic cavitation is a dominant factor in both stone comminution and kidney damage. A competing mechanism is the shear stress produced in tissue by the SWL shock wave. Acoustic cavitation results from the growth and violent collapse of cavitation bubbles produced by the SWL acoustic waveform. We have developed a set of sophisticated experimental tools that permit us to detect cavitation in a broad range of environments and with some degree of spatial and temporal resolution. Using these tools, we have acquired a significant body of evidence that demonstrates that SWL generates cavitation in both the parenchyma and the collecting system of the human kidney. We have also discovered a method of modifying an electrohydraulic lithotripter to produce a waveform that has a similar shock wave amplitude, pulse length, and acoustic energy to that of a conventional SWL waveform, but does not generate cavitation. Furthermore, we have found a way to enhance the violence cavitation collapse, as well as a way to confine cavitation to a highly localized volume. We propose to use these various discoveries and tools to ascertain to ascertain the relative roles of cavitation and shear in stone comminution and tissue damage. We have also made considerable progress toward the development of a set of theoretical models that would permit us to compute the SWL waveform at any position and time, either in vitro or in vivo. Furthermore, give a specific waveform, we can compute the response of the cavitation field to this waveform, and determine, albeit in a crude way, the potential for cavitation damage in a variety of in vitro and in vivo environments. We propose to test our hypotheses by undertaking a series of experiments in which we will apply SWL to various in vitro and in vivo models, during which we will use our cavitation detection techniques to determine the presence and location of this cavitation; later, we shall correlate the tissue damage with either the presence or absence of cavitation. We shall also undertake a series of similar experiments involving stone comminution. Using these results, we shall use our theoretical models to design a waveform that would optimize stone comminution and minimize tissue damage.

Project Start
1999-04-01
Project End
2000-02-29
Budget Start
1998-10-01
Budget End
1999-09-30
Support Year
6
Fiscal Year
1999
Total Cost
Indirect Cost
Name
Indiana University-Purdue University at Indianapolis
Department
Type
DUNS #
005436803
City
Indianapolis
State
IN
Country
United States
Zip Code
46202
Janssen, Karmon M; Brand, Timothy C; Bailey, Michael R et al. (2018) Effect of Stone Size and Composition on Ultrasonic Propulsion Ex Vivo. Urology 111:225-229
Simon, Julianna C; Sapozhnikov, Oleg A; Kreider, Wayne et al. (2018) The role of trapped bubbles in kidney stone detection with the color Doppler ultrasound twinkling artifact. Phys Med Biol 63:025011
Matula, Thomas J; Sapozhnikov, Oleg A; Ostrovsky, Lev A et al. (2018) Ultrasound-based cell sorting with microbubbles: A feasibility study. J Acoust Soc Am 144:41
Williams Jr, James C; Borofsky, Michael S; Bledsoe, Sharon B et al. (2018) Papillary Ductal Plugging is a Mechanism for Early Stone Retention in Brushite Stone Disease. J Urol 199:186-192
Sapozhnikov, Oleg; Nikolaeva, Anastasiia; Bailey, Michael (2018) The effect of shear waves in an elastic sphere on the radiation force from a quasi-Gaussian beam. Proc Meet Acoust 32:
Zwaschka, Theresa A; Ahn, Justin S; Cunitz, Bryan W et al. (2018) Combined Burst Wave Lithotripsy and Ultrasonic Propulsion for Improved Urinary Stone Fragmentation. J Endourol 32:344-349
Connors, Bret A; Schaefer, Ray B; Gallagher, John J et al. (2018) Preliminary Report on Stone Breakage and Lesion Size Produced by a New Extracorporeal Electrohydraulic (Sparker Array) Discharge Device. Urology 116:213-217
Dai, Jessica C; Dunmire, Barbrina; Sternberg, Kevan M et al. (2018) Retrospective comparison of measured stone size and posterior acoustic shadow width in clinical ultrasound images. World J Urol 36:727-732
Lingeman, James E (2017) Editorial Comment. J Urol 197:1572
Cunitz, Bryan W; Dunmire, Barbrina; Bailey, Michael R (2017) Characterizing the Acoustic Output of an Ultrasonic Propulsion Device for Urinary Stones. IEEE Trans Ultrason Ferroelectr Freq Control 64:1818-1827