The overall goal of this renewal application is to improve the effectiveness of single photon emission tomography (SPECT) imaging for the diagnosis and treatment of patients with cancer. More specifically, it focuses on the capabilities of modern SPECT systems that include X-ray computed tomography systems (CT) with a focus on the estimation of activity concentration. This proposed effort will include a systematic phantom exploration of two SPECT/CT systems (one with cone beam CT and one with 16 slice helical CT) with 16 different combinations of isotopes and collimators. The data will be collected in the current sinogram binned mode and, in parallel, in listmode (keeping all events above 50 keV). A 4 cm diameter 57Co cylinder with calibrated 57Co standards for the dose calibrator will be used to determine a conversion factor from counts/pixel to activity/volume. These data will be used to explore the feasibility of such a calibration measurement (one isotope, one collimator) and scaling factors to apply that calibration to all isotope/collimator combinations. As part of ths effort we will develop more accurate scatter and collimator response modeling using accelerated Monte Carlo simulation in the image reconstruction process. This new effort is an outgrowth of our previous work, under this grant funding, in exploring quantitation in PET, including isolating many sources of error, finding ways to minimize them, and exploring alternative scanner geometries. That work also included image reconstruction with modeling scanner non-stationary resolution response. As part of our previously funded research, a Simulation System for Emission Tomography (SimSET) has been developed. This Monte Carlo-based software package tracks photons through complex heterogeneous objects (patients) and models the effects of collimation and detection. The package has been essential to our efforts and has been adopted by more than 300 other investigators in their own work. Part of our effort will be to continue to improve SimSET with significance acceleration of SPECT collimator simulation tools. The SimSET software package is freely available via the Internet.

Public Health Relevance

Single Photon emission tomography (SPECT) has now been combined with X-ray Computed tomography (CT) to provide a single scanner that can image both physiology and anatomy. This grant will develop techniques to make quantitative SPECT/CT more available as a practical and cost effective clinical tool to make more specific diagnosis of disease and monitoring response to therapy than current techniques allow. This work is another step toward personalized medicine where therapies are tailored to the individual patient.

Agency
National Institute of Health (NIH)
Institute
National Cancer Institute (NCI)
Type
Research Project (R01)
Project #
5R01CA042593-26
Application #
8677694
Study Section
Special Emphasis Panel (ZRG1-BMIT-J (01))
Program Officer
Zhang, Yantian
Project Start
1987-07-01
Project End
2017-05-31
Budget Start
2014-06-01
Budget End
2015-05-31
Support Year
26
Fiscal Year
2014
Total Cost
$322,162
Indirect Cost
$113,643
Name
University of Washington
Department
Radiation-Diagnostic/Oncology
Type
Schools of Medicine
DUNS #
605799469
City
Seattle
State
WA
Country
United States
Zip Code
98195
Zuki?, Dženan; Byrd, Darrin W; Kinahan, Paul E et al. (2018) Calibration Software for Quantitative PET/CT Imaging Using Pocket Phantoms. Tomography 4:148-158
Kinahan, Paul E; Byrd, Darrin W; Helba, Brian et al. (2018) Simultaneous Estimation of Bias and Resolution in PET Images With a Long-Lived ""Pocket"" Phantom System. Tomography 4:33-41
Zeng, Chengeng; Kinahan, Paul E; Qian, Hua et al. (2017) Simulation study of quantitative precision of the PET/X dedicated breast PET scanner. J Med Imaging (Bellingham) 4:045502
Lee, Tzu C; Alessio, Adam M; Miyaoka, Robert M et al. (2016) Morphology supporting function: attenuation correction for SPECT/CT, PET/CT, and PET/MR imaging. Q J Nucl Med Mol Imaging 60:25-39
Gong, Kuang; Majewski, Stan; Kinahan, Paul E et al. (2016) Designing a compact high performance brain PET scanner-simulation study. Phys Med Biol 61:3681-97
McDougald, Wendy A; Miyaoka, Robert S; Alessio, Adam M et al. (2016) A study of SPECT/CT camera stability for quantitative imaging. EJNMMI Phys 3:14
Vaquero, Juan José; Kinahan, Paul (2015) Positron Emission Tomography: Current Challenges and Opportunities for Technological Advances in Clinical and Preclinical Imaging Systems. Annu Rev Biomed Eng 17:385-414
Harrison, Robert L; Elston, Brian F; Doot, Robert K et al. (2014) A Virtual Clinical Trial of FDG-PET Imaging of Breast Cancer: Effect of Variability on Response Assessment. Transl Oncol 7:138-46
MacDonald, L R; Harrison, R L; Alessio, A M et al. (2011) Effective count rates for PET scanners with reduced and extended axial field of view. Phys Med Biol 56:3629-43
Harrison, Robert L (2010) Introduction To Monte Carlo Simulation. AIP Conf Proc 1204:17-21