It is proposed a diagnostic method for the on-line, semi- continuous determination of the size and density of individual particles in the 5 nm to 100 nm size range. This is accomplished through a double size spectrometry system. Particles are first sized by their Stokes diameter, and then sized by their aerodynamic diameter. The Stokes diameter sizing gives the geometric diameter of the particle for round particles. The aerodynamic sizing is dependent on particle density as well as Stokes diameter. The result is a distribution of particle densities at each selected Stokes diameter. The Stokes sizing is done by an electrical mobility separation, the aerodynamic sizing is done using a single-stage hypersonic impactor. Since particles are singly charged, the rate of particle capture in can be measured on-line using an electrometer. The system is effective over the 5 to 100 nm diameter range.

Agency
National Science Foundation (NSF)
Institute
Division of Industrial Innovation and Partnerships (IIP)
Type
Standard Grant (Standard)
Application #
9061198
Program Officer
Ritchie B. Coryell
Project Start
Project End
Budget Start
1991-01-01
Budget End
1991-09-30
Support Year
Fiscal Year
1990
Total Cost
$49,852
Indirect Cost
Name
Aerosol Dynamics
Department
Type
DUNS #
City
Berkeley
State
CA
Country
United States
Zip Code
94710