The goal of this project is to develop new radiofrequency (RF) gradient encoding methods to spatially encode signals in MRI. The methods would enable silent, low-cost MRI systems, leading to a substantial reduction in the cost of imaging and improved patient compliance and comfort. In conventional MRI, a received signal is localized to its spatial location of origin based on its temporal frequency, which is controlled using magnetic fields that are parallel to the main (B0) field of the scanner and vary linearly across space. There are many problems with these B0 gradient fields: they are loud and induce peripheral nerve stimulation, compromising patient comfort;they have relatively long switching times due to the high inductance of the coils;they require bulky cooling systems and customized amplifiers;and they are expensive, representing around 20-25% of the cost of a clinical scanner. A potential solution to these problems is to replace B0 gradients with RF gradients, which are silent and low-cost. Unfortunately, in spite of its potential RF gradient encoding has not yet become a clinical or commercial success. This is largely due to the fact that no existing RF gradient encoding method offers the orthogonality between contrast development and spatial encoding that is en- joyed by B0 gradients, or a straightforward path to convert existing B0 gradient-based MRI acquisition techniques to use RF encoding. The methods proposed in this project will be the first to meet these requirements, and will thus represent the first truly viable RF gradient-based imaging methods. The central innovation of this project is to use the Bloch-Siegert shift to spatially encode the MRI signal. As with B0 gradients, this encoding mechanism is based on the application of phase shifts to magnetization directly in the transverse plane, and therefore does not modulate the magnitude of the transverse magnetization, leaving image contrast unaffected by spatial encoding.
The first Aim of the project is to develop new RF gradient coils and other RF hardware to enable 2D and 3D Cartesian imaging at 0.5 Tesla, including hardware strategies for simultaneous RF transmission and reception to enable frequency encoding by Bloch-Siegert shift.
The second Aim i s to develop new RF-encoded pulse sequences based on the Bloch-Siegert shift, leveraging recent innovations in RF pulse design for Bloch-Siegert phase encoding, and in RF pulse design for RF gradient-based slice-selective excitation.
The third Aim i s to develop robust algorithms to reconstruct images from RF-encoded data. Successful completion of these Aims would broadly prove the feasibility of the proposed RF encoding methods for MR imaging, paving the way for translation to humans.

Public Health Relevance

The hardware currently used for spatial encoding in MRI causes significant patient discomfort due to noise and peripheral nerve stimulation, and represents a large portion of the cost of a scanner. This project aims to address these problems through the development of new spatial encoding strategies using radiofrequency fields. The successful completion of this project has the potential to bring about a new generation of silent, more compact and lower-cost MRI scanners.

Agency
National Institute of Health (NIH)
Institute
National Institute of Biomedical Imaging and Bioengineering (NIBIB)
Type
Exploratory/Developmental Grants (R21)
Project #
1R21EB018521-01A1
Application #
8828416
Study Section
Biomedical Imaging Technology Study Section (BMIT)
Program Officer
Liu, Guoying
Project Start
2014-09-25
Project End
2016-07-31
Budget Start
2014-09-25
Budget End
2015-07-31
Support Year
1
Fiscal Year
2014
Total Cost
$189,980
Indirect Cost
$64,980
Name
Vanderbilt University Medical Center
Department
Biomedical Engineering
Type
Schools of Engineering
DUNS #
004413456
City
Nashville
State
TN
Country
United States
Zip Code
37212
Yan, Xinqiang; Gore, John C; Grissom, William A (2018) Self-decoupled radiofrequency coils for magnetic resonance imaging. Nat Commun 9:3481
Yan, Xinqiang; Cao, Zhipeng; Grissom, William A (2018) Ratio-adjustable power splitters for array-compressed parallel transmission. Magn Reson Med 79:2422-2431
Yan, Xinqiang; Gore, John C; Grissom, William A (2017) New resonator geometries for ICE decoupling of loop arrays. J Magn Reson 277:59-67
Yan, Xinqiang; Zhang, Xiaoliang; Gore, John C et al. (2017) Improved traveling-wave efficiency in 7T human MRI using passive local loop and dipole arrays. Magn Reson Imaging 39:103-109
Hasselwander, Christopher J; Cao, Zhipeng; Grissom, William A (2016) gr-MRI: A software package for magnetic resonance imaging using software defined radios. J Magn Reson 270:47-55