Continuous glucose monitoring (CGM) involves repetitive measurement of physiological glucose concentration to enable close monitoring and timely correction of problematic blood sugar patterns of patients with diabetes mellitus, thereby reducing the risk of diabetes-related complications and ultimately allowing closed-loop blood sugar monitoring and insulin administration. Commercially available CGM sensors that use electrochemical methods are currently hindered by limitations such as low accuracy (especially at hypoglycemic glucose concentrations), poor stability, and long lag times.
We aim to address these issues by developing a subcutaneously implantable affinity microsensor for continuous monitoring of glucose in interstitial fluid. The microsensor, created using microelectromechanical systems (MEMS) technology, will have a miniaturized, flexible design to realize differential measurement of affinity binding of glucose to a synthetic, biocompatible hydrogel. Affinity binding, in which glucose binds specifically and reversibly to the hydrogel without glucose- consuming chemical reactions commonly found in existing glucose sensors, affords high stability. MEMS- based differential measurement of affinity binding enables rapid, accurate determination of glucose concentration, in particular in the hypoglycemic range, in the face of nonspecific disturbances. These functions are realized in a miniaturized, flexible design, which minimizes the effects of device-tissue interactions. In design, the microsensor resides on a flexible substrate and is integrated with a glucose-binding (sensing) hydrogel and a glucose-insensitive (reference) hydrogel. With an active sensing region hundreds of micrometers in size, the device is implanted (via a small needle) beneath the skin in the abdominal region. During operation, glucose molecules in tissue rapidly enter the microsensor and bind reversibly to the sensing hydrogel, whose dielectric properties change accordingly. Meanwhile, the reference hydrogel's dielectric properties change only with nonspecific disturbances (e.g., temperature). Thus, differential dielectric measurement allows accurate determination of the glucose concentration in the interstitial fluid. The direct goal of this project is to develop the differential affinity microsensor for percutaneously implanted operation over a period of 5-7 days with a high level of stability and accuracy.
The specific aims i nclude (1) functional hydrogel synthesis, (2) device design and fabrication, and (3) hydrogel and device characterization in vitro and in vivo. The device will in the future be further developed to allow long-term (months or longer) implanted operation, and be included in an artificial pancreas to enable closed-loop glucose control.

Public Health Relevance

Currently, approximately 25.8 million people in the U.S (8.3 percent of the population) have diabetes. Close monitoring of treatment with repetitive daily blood glucose measurements has been demonstrated to reduce the risk of diabetes-related complications. Thus, the development of devices for reliable and accurate glucose monitoring, as pursued in this project, is highly relevant to public health.

Agency
National Institute of Health (NIH)
Institute
National Institute of Diabetes and Digestive and Kidney Diseases (NIDDK)
Type
Type 1 Diabetes Targeted Research Award (DP3)
Project #
1DP3DK101085-01
Application #
8642995
Study Section
Special Emphasis Panel (ZDK1-GRB-N (O1))
Program Officer
Arreaza-Rubin, Guillermo
Project Start
2013-09-30
Project End
2016-06-30
Budget Start
2013-09-30
Budget End
2016-06-30
Support Year
1
Fiscal Year
2013
Total Cost
$1,100,000
Indirect Cost
$308,070
Name
Columbia University (N.Y.)
Department
Engineering (All Types)
Type
Schools of Engineering
DUNS #
049179401
City
New York
State
NY
Country
United States
Zip Code
10027
Rushing, J; Wing, R; Wadden, T A et al. (2017) Cost of intervention delivery in a lifestyle weight loss trial in type 2 diabetes: results from the Look AHEAD clinical trial. Obes Sci Pract 3:15-24
Shang, Junyi; Yan, Jing; Zhang, Zhixing et al. (2016) A hydrogel-based glucose affinity microsensor. Sens Actuators B Chem 237:992-998
Zhu, Yibo; Hao, Yufeng; Adogla, Enoch A et al. (2016) A graphene-based affinity nanosensor for detection of low-charge and low-molecular-weight molecules. Nanoscale 8:5815-9
Wang, Cheng; Kim, Jinho; Zhu, Yibo et al. (2015) An aptameric graphene nanosensor for label-free detection of small-molecule biomarkers. Biosens Bioelectron 71:222-229
Huang, Xian; Leduc, Charles; Ravussin, Yann et al. (2014) A differential dielectric affinity glucose sensor. Lab Chip 14:294-301