Sudden cardiac death is a major cause of mortality in the United States. In an effort to attach this problem, the concept of a Chain of Survival has been presented as a way to characterized the tasks necessary for successful resuscitation. It is clear that early defibrillation is central to improved survival rates following out-of-hospital cardiac arrest. 1) successful defibrillation of the arrhythmia of the arrhythmia, 2) recurrence if arrhythmias following defibrillation, and 3) poor cardiac function following defibrillation This project will address the first two problems. Very little is known about defibrillation efficacy of spontaneous arrhythmias in the presence of acute ischemia. A much larger shock is needed to halt ventricular fibrillation that occurs spontaneously in the presence of acute is ischemia than to halt electrically induced ventricular fibrillation in the non-ischemic heart. We will study the activation patter following successful and failed defibrillation shocks with the heart in four conditions designed to model different aspects of ischemic ventricular fibrillation shocks the heart in four conditions designed to model different aspects of ischemic ventricular fibrillation: (1) acute regional ischemia causing spontaneous ventricular fibrillation, (2) acute regional ischemia followed by electrically induced ventricular fibrillation, (3) acute regional ischemia causing spontaneous ventricular fibrillation which in turn causes prolonged global ischemia, and (4) acute regional ischemia causing spontaneous ventricular fibrillation in the setting of prolonged global ischemia, and (4) acute regional ischemia causing spontaneous ventricular fibrillation in the setting of an old myocardial infarction. We will test the hypothesis that, in the setting of an ischemically induced spontaneous arrhythmia, a defibrillation shock must do three things: (1) stop all fibrillation wavefronts, (2) not restart fibrillation, and (3) stop the trigger of the original arrhythmia. We will test the hypothesis that a major determinant of whether or not a spontaneous arrhythmia, a defibrillation shock must do three things: (1) stop all fibrillation wavefronts, (2)m not restart fibrillation and (3) stop the trigger of the original. We will test the hypothesis that a major determinant f whether or not a spontaneous arrhythmia caused by acute ischemia can be easily defibrillated depends upon the mechanism of initiation of the arrhythmia and duration of the arrhythmia. We will also test the hypothesis that arrhythmias that occur following the initial fibrillation episode are initiated by focal rather than reentrant mechanisms and will be harder to defibrillate than the initial arrhythmia. To test these hypothesis, we will induce acute ischemia by creating a thrombus in a coronary artery and use electrical mapping techniques to map the initiation of ventricular arrhythmias, and the first few post-shock activations following delivery of a defibrillation shock. By developing an understanding of how a defibrillation shock succeeds or fails under these conditions, we will be in a better position to develop new defibrillation techniques that increase survival rates of patients suffering an episode of sudden cardiac death.

Agency
National Institute of Health (NIH)
Institute
National Heart, Lung, and Blood Institute (NHLBI)
Type
Research Program Projects (P01)
Project #
5P01HL067961-02
Application #
6630622
Study Section
Heart, Lung, and Blood Initial Review Group (HLBP)
Project Start
2002-08-01
Project End
2003-07-31
Budget Start
2002-08-01
Budget End
2003-07-31
Support Year
2
Fiscal Year
2002
Total Cost
$164,585
Indirect Cost
Name
University of Alabama Birmingham
Department
Type
DUNS #
004514360
City
Birmingham
State
AL
Country
United States
Zip Code
35294
Walcott, Gregory P; Melnick, Sharon B; Killingsworth, Cheryl R et al. (2010) Comparison of low-energy versus high-energy biphasic defibrillation shocks following prolonged ventricular fibrillation. Prehosp Emerg Care 14:62-70
Walcott, Gregory; Melnick, Sharon; Killingsworth, Cheryl et al. (2009) Burst stimulation improves hemodynamics during resuscitation after prolonged ventricular fibrillation. Circ Arrhythm Electrophysiol 2:57-62
Doppalapudi, Harish; Jin, Qi; Dosdall, Derek J et al. (2008) Intracoronary infusion of catecholamines causes focal arrhythmias in pigs. J Cardiovasc Electrophysiol 19:963-70
Dosdall, Derek J; Cheng, Kang-An; Huang, Jian et al. (2007) Transmural and endocardial Purkinje activation in pigs before local myocardial activation after defibrillation shocks. Heart Rhythm 4:758-65
Dosdall, Derek J; Ideker, Raymond E (2007) Intracardiac atrial defibrillation. Heart Rhythm 4:S51-6
Kong, Wei; Huang, Jian; Rollins, Dennis L et al. (2007) A semi-implantable multichannel telemetry system for continuous electrical, mechanical and hemodynamical recordings in animal cardiac research. Physiol Meas 28:249-57
Lan, David Z; Pollard, Andrew E; Knisley, Stephen B et al. (2007) Optical mapping of V(m) and Ca(i)(2+) in a model of arrhythmias induced by local catecholamine application in patterned cell cultures. Pflugers Arch 453:871-7
Raman, Vidya; Pollard, Andrew E; Fast, Vladimir G (2007) Shock-induced changes of Ca(i)2+ and Vm in myocyte cultures and computer model: Dependence on the timing of shock application. Cardiovasc Res 73:101-10
Pollard, Andrew E; Barr, Roger C (2006) Multisite interstitial stimulation for cardiac micro-impedance measurements. Conf Proc IEEE Eng Med Biol Soc 1:1572-5
Dosdall, Derek J; Huang, Jian; Smith, William M et al. (2006) Printed circuit board electrodes for transmural cardiac mapping. Conf Proc IEEE Eng Med Biol Soc 1:3927-30

Showing the most recent 10 out of 35 publications