Magnetic Resonance Imaging (MRI) has become the gold standard in the imaging of many soft tissues, osseous and joint conditions, but to date there has been limited success in harnessing this excellent imaging modality for interventional procedures. At the same time, MRI poses formidable research challenges by limited access to the patient and a strong magnetic field that prevents the use of conventional materials and electronic equipment. Currently, no exigent technical solution exists to assist MRI-guided needle placement procedures in an accurate, simple, and economical manner. The objective of the proposed research program is to make diagnostic closed high-field MRI scanners available for guiding needle placement in musculoskeletal and spinal interventions. Our approach is to apply image overlay technique to make an MR image virtually float inside the patient in correct alignment and magnification, as if the operator had """"""""MRI vision"""""""" to virtually slice through the body. The technical hypothesis is that MR image overlay guidance is suitable for a variety of needle-based interventional procedures and it allows for appropriate procedural accuracy and consistency in musculoskeletal and spinal interventions on conventional closed high-field MRI scanners.
The specific aims toward proving our technical hypothesis are the following: (1) Develop a clinically adequate and integrated image overlay system to guide needle insertion on high-field (1.5T-3T) closed MRI scanners. (2) Refine procedural workflow and perform technical validation of the system on phantoms and human cadavers. (3) Perform clinical system performance evaluation trial on human subjects in joint arthrography, MSK biopsy (sarcoma), vertebral biopsy, and disk space aspiration procedures. The potential societal benefit of the proposed research is significant. Our long-term objective is to produce generally applicable technology for in-room MRI-guided needle placement. As MRI is becoming gradually affordable, the MR image overlay technique may facilitate a wide array of diagnostic and therapeutic interventions involving percutaneous needle insertion. An effective, simple, and affordable guidance system would provide greater access for facilities using the current install base of magnets. It can also reduce variability among practitioners, facilitate training, and serve as a research validation tool by providing precisely placed minimally invasive percutaneous conduit under MR guidance.

Agency
National Institute of Health (NIH)
Institute
National Cancer Institute (NCI)
Type
Research Project (R01)
Project #
5R01CA118371-03
Application #
7666068
Study Section
Biomedical Imaging Technology Study Section (BMIT)
Program Officer
Farahani, Keyvan
Project Start
2007-09-05
Project End
2011-07-31
Budget Start
2009-08-01
Budget End
2010-07-31
Support Year
3
Fiscal Year
2009
Total Cost
$340,594
Indirect Cost
Name
Johns Hopkins University
Department
Radiation-Diagnostic/Oncology
Type
Schools of Medicine
DUNS #
001910777
City
Baltimore
State
MD
Country
United States
Zip Code
21218
Marker, David R; U Thainual, Paweena; Ungi, Tamas et al. (2017) 1.5 T augmented reality navigated interventional MRI: paravertebral sympathetic plexus injections. Diagn Interv Radiol 23:227-232
Marker, David R; U-Thainual, Paweena; Ungi, Tamas et al. (2016) MR-guided perineural injection of the ganglion impar: technical considerations and feasibility. Skeletal Radiol 45:591-7
Anand, M; King, F; Ungi, T et al. (2014) Design and development of a mobile image overlay system for needle interventions. Conf Proc IEEE Eng Med Biol Soc 2014:6159-62
Fritz, Jan; U-Thainual, Paweena; Ungi, Tamas et al. (2014) MR-guided vertebroplasty with augmented reality image overlay navigation. Cardiovasc Intervent Radiol 37:1589-96
Fritz, Jan; U-Thainual, Paweena; Ungi, Tamas et al. (2013) Augmented reality visualization using image overlay technology for MR-guided interventions: cadaveric bone biopsy at 1.5 T. Invest Radiol 48:464-70
Fritz, Jan; U-Thainual, Paweena; Ungi, Tamas et al. (2013) Augmented reality visualisation using an image overlay system for MR-guided interventions: technical performance of spine injection procedures in human cadavers at 1.5 Tesla. Eur Radiol 23:235-45
U-Thainual, Paweena; Fritz, Jan; Moonjaita, Choladawan et al. (2013) MR image overlay guidance: system evaluation for preclinical use. Int J Comput Assist Radiol Surg 8:365-78
Fritz, Jan; U-Thainual, Paweena; Ungi, Tamas et al. (2012) Augmented reality visualization with image overlay for MRI-guided intervention: accuracy for lumbar spinal procedures with a 1.5-T MRI system. AJR Am J Roentgenol 198:W266-73
Fayad, Laura M; Jacobs, Michael A; Wang, Xin et al. (2012) Musculoskeletal tumors: how to use anatomic, functional, and metabolic MR techniques. Radiology 265:340-56
Ahmidi, Narges; Ishii, Masaru; Fichtinger, Gabor et al. (2012) An objective and automated method for assessing surgical skill in endoscopic sinus surgery using eye-tracking and tool-motion data. Int Forum Allergy Rhinol 2:507-15

Showing the most recent 10 out of 13 publications