Breast cancer remains an important disease, adversely affecting a large population of women. The number of women diagnosed with breast cancer has continued to rise. While the majority of patients survive the disease, minimizing any disability or disfigurement as a result of treatment is very important to the patient's future qualty of life. Despite more sensitive MRI methods to detect breast lesions and new minimally invasive conservative forms of therapy, there still remains significant room for improvement in both imaging and therapy. MRI-guided high intensity focused ultrasound (MRgHIFU) has the potential to provide completely non-invasive therapy of localized lesions in the breast. This proposal addresses the remaining limitations that keep MRgHIFU from achieving its potential to non-invasively treat localized breast cancer. Building on an innovative breast-specific MRgHIFU system developed in an academic/industrial partnership led by scientists in the Utah Center for Advanced Imaging Research, the College of Engineering, and School of Medicine, this project will validate the system for rapid, efficacious, and safe non-invasive treatment of localized breast lesions with MRgHIFU. The project consists of four Specific Aims: 1) develop patient-specific methods of ultrasound beam aberration correction to improve ultrasound focus localization in the heterogeneous breast; 2) develop efficient methods for temperature measurement in the adipose and aqueous tissues of the heterogeneous breast to properly monitor the MRgHIFU treatment procedure thereby ensuring complete tumor ablation and avoiding unwanted tissue damage; 3) demonstrate safety and efficacy of tumor and treatment margin ablation in an animal model by proving the system can achieve complete ablation in 90% of all subjects without violating safety constraints; and 4) evaluate the developed ultrasound beam aberration correction and temperature measurement techniques in vivo in humans to demonstrate that the developed techniques are effective in human breasts. The innovations proposed in this project are critical for the safe non-invasive treatment of breast cancer with MRgHIFU and will lay the foundation for a successful phase I clinical trial. This new technology will provide a new, non-invasive treatment option for women with localized breast cancer.

Public Health Relevance

This project will develop missing crucial capabilities to enable the application of MRI-guided high intensity focused ultrasound therapy to the heterogeneous human breast and perform preclinical studies of system safety and efficacy. This new technology will provide a new, non-invasive treatment option for women with localized breast cancer.

Agency
National Institute of Health (NIH)
Institute
National Cancer Institute (NCI)
Type
Research Project (R01)
Project #
5R01CA172787-03
Application #
8858591
Study Section
Biomedical Imaging Technology Study Section (BMIT)
Program Officer
Farahani, Keyvan
Project Start
2013-08-21
Project End
2016-05-31
Budget Start
2015-06-01
Budget End
2016-05-31
Support Year
3
Fiscal Year
2015
Total Cost
Indirect Cost
Name
University of Utah
Department
Radiation-Diagnostic/Oncology
Type
Schools of Medicine
DUNS #
009095365
City
Salt Lake City
State
UT
Country
United States
Zip Code
84112
de Bever, Joshua T; Odéen, Henrik; Hofstetter, Lorne W et al. (2018) Simultaneous MR thermometry and acoustic radiation force imaging using interleaved acquisition. Magn Reson Med 79:1515-1524
Freeman, Nicholas J; Odéen, Henrik; Parker, Dennis L (2018) 3D-specific absorption rate estimation from high-intensity focused ultrasound sonications using the Green's function heat kernel. Med Phys 45:3109-3119
Svedin, Bryant T; Beck, Michael J; Hadley, J Rock et al. (2017) Focal point determination in magnetic resonance-guided focused ultrasound using tracking coils. Magn Reson Med 77:2424-2430
Svedin, Bryant T; Parker, Dennis L (2017) Technical Note: The effect of 2D excitation profile on T1 measurement accuracy using the variable flip angle method with an average flip angle assumption. Med Phys 44:5930-5937
Shi, Y C; Parker, D L; Dillon, C R (2016) Sensitivity of tissue properties derived from MRgFUS temperature data to input errors and data inclusion criteria: ex vivo study in porcine muscle. Phys Med Biol 61:N373-85
Farrer, Alexis I; Almquist, Scott; Dillon, Christopher R et al. (2016) Phase aberration simulation study of MRgFUS breast treatments. Med Phys 43:1374-84
Johnson, Sara L; Dillon, Christopher; Odéen, Henrik et al. (2016) Development and validation of a MRgHIFU non-invasive tissue acoustic property estimation technique. Int J Hyperthermia 32:723-34
Svedin, Bryant T; Payne, Allison; Parker, Dennis L (2016) Respiration artifact correction in three-dimensional proton resonance frequency MR thermometry using phase navigators. Magn Reson Med 76:206-13
de Bever, Joshua T; Odéen, Henrik; Todd, Nick et al. (2016) Evaluation of a three-dimensional MR acoustic radiation force imaging pulse sequence using a novel unbalanced bipolar motion encoding gradient. Magn Reson Med 76:803-13
Almquist, Scott; Parker, Dennis L; Christensen, Douglas A (2016) Rapid full-wave phase aberration correction method for transcranial high-intensity focused ultrasound therapies. J Ther Ultrasound 4:30

Showing the most recent 10 out of 36 publications