Most presently installed full-body MRI systems are based on 1.5 T liquid-He- cooled NbTi magnets, but the rate of installation of 3.0 T systems is rapidly increasing. However, the increasing cost of helium is well documented, and this leads to increases in the life cycle cost of MRI systems, which could limit ready access for traditional patient groups, and limit the ability to expand availability to traditionally less well served patint groups. The root cause of the problem is the growing shortage of helium worldwide. The major MRI magnet manufacturers (Siemens, GE, and Phillips) each have development programs focused on converting the 1.5T solenoid coils from liquid helium bath cooling to dry (cryogen free) conduction cooling. Indeed, freedom from the need for liquid-helium cooling (i.e. cryogen free, conduction-cooled, operation) is becoming more and more important. Presently, MRI systems (2000-3000 liters) are filled with liquid He helium at the factory prior to shipping at a cost in the U.S. of about 20-30 K. Given that from 2006 to 2011 the cost of LHe has tripled, and it is expected to continue to increase, corresponding increases in the cost of liquid-He MRI units can be expected. In some parts of the world liquid He costs about $9-13 per liter and in others about $25 to $70 per liter leading to re-filling costs of $18,000-$210,000 depending on the MRI's size and location. Thus a growing demand is predicted for MRI magnets that do not depend on liquid He for cooling. Even beyond the liquid cryogen issues for MRI, the development of cryocooled MRI magnets could allow for new geometries and MRI applications, opening up new treatment possibilities. However, we must focus where the industry can transition this new technology commercially. Accordingly, the proposed program focuses on developing the technology to enable cryogen free 3 T systems using a cryocooled mode. We proposed to do this by; (1) Increases practical MgB2 engineering current density, Je, by one order of magnitude over presently developed 1 G wire, in the fields and temperature ranges appropriate to 3 T MRI (4-10 K, 0-4 T), (2) Developing n-values sufficient for persistent mode magnet operation (30 or more in the relevant field range) in the multifilament 2 G wires of (1), (3) Developing a magnet model with sufficient detail to develop the conductor specification/targets, and the tradeoffs between cooling, temperature gradients across a coil segment, and conductor performance, (4) Determining the proper Cu/SC ratio for wire-in-channel MgB2/Cu, and demonstrate its long-length manufacture, (5) Demonstrating one segment of the total coil, using the WIC of item (4), with the needed current, n-value, temperature gradient, and quench performance, and (6) Developing a robust persistent joint for MgB2 conductors in a magnet engineering friendly react & wind mode.

Public Health Relevance

Present day MRI systems, whether 1.5 T or 3 T, use NbTi superconductors and operate in liquid helium. However, the cost of liquid helium has been rising rapidly, and is expected to continue to do so, helium is a finite resource. Thus, costs for systems based on liquid helium are expected to increase, with potential negative impact on MRI availability, or machine and thus cost-per scan. This proposed program aims to develop essential technology to permit systems which can operate with much lower amounts of helium using a cryocooling method. This will also enable new MRI applications not constrained by the designs requiring liquid cryogen cooling, thus opening up new treatment possibilities. In order to do this, improvements in conductor, magnet design, and wire joints are needed. However, program success would lead to reduced costs and improved health care for the public.

Agency
National Institute of Health (NIH)
Institute
National Institute of Biomedical Imaging and Bioengineering (NIBIB)
Type
Research Project (R01)
Project #
5R01EB018363-04
Application #
9418597
Study Section
Biomedical Imaging Technology Study Section (BMIT)
Program Officer
Wang, Shumin
Project Start
2015-01-01
Project End
2019-12-31
Budget Start
2018-01-01
Budget End
2019-12-31
Support Year
4
Fiscal Year
2018
Total Cost
Indirect Cost
Name
Ohio State University
Department
Engineering (All Types)
Type
Biomed Engr/Col Engr/Engr Sta
DUNS #
832127323
City
Columbus
State
OH
Country
United States
Zip Code
43210
Wan, F; Sumption, M D; Rindfleisch, M A et al. (2017) Architecture and Transport Properties of Multifilamentary MgB2 Strands for MRI and Low AC Loss Applications. IEEE Trans Appl Supercond 27:
Parizh, Michael; Lvovsky, Yuri; Sumption, Michael (2017) Conductors for commercial MRI magnets beyond NbTi: requirements and challenges. Supercond Sci Technol 30:014007
Baig, Tanvir; Al Amin, Abdullah; Deissler, Robert J et al. (2017) Conceptual designs of conduction cooled MgB2 magnets for 1.5 and 3.0T full body MRI systems. Supercond Sci Technol 30:
Kim, H S; Kovacs, C; Rindfleisch, M et al. (2016) Demonstration of a Conduction Cooled React and Wind MgB2 Coil Segment for MRI Applications. IEEE Trans Appl Supercond 26:
Yang, Y; Sumption, M D; Collings, E W (2016) Influence of Metal Diboride and Dy2O3 Additions on Microstructure and Properties of MgB2 Fabricated at High Temperatures and under Pressure. Sci Rep 6:29306
Li, G Z; Sumption, M D; Collings, E W (2015) Kinetic analysis of MgB2 layer formation in advanced internal magnesium infiltration (AIMI) processed MgB2 wires. Acta Mater 96:66-71
Yang, Y; Li, G; Susner, M et al. (2015) Influence of Twisting and Bending on the Jc and n-value of Multifilamentary MgB2 Strands. Physica C Supercond 519:118-123
Li, G Z; Susner, M A; Bohnenstiehl, S D et al. (2015) Microstructures and superconducting properties of high performance MgB2 thin films deposited from a high-purity, dense Mg-B target. Appl Surf Sci 357 Pt A:452-458