We propose to develop a miniature real-time volumetric ultrasound imaging system employing two-dimensional (2D) capacitive micromachined ultrasonic transducer (CMUT) arrays. We will design and implement a miniature probe fitting into the tip of an endoscope that will be connected to an external beamforming unit using a small number of cables. The probe will contain a 2D CMUT array and integrated circuits to handle the data acquisition and front-end pre-processing. A small-sized active phased subarray will be multiplexed over a large transducer array in consecutive firing steps to allow for real-time data acquisition. The acquired data will be transferred to the external unit and processed to form real-time pyramidal volumetric images. CMUT technology provides excellent element sensitivity, large temporal bandwidth, and potential for integrating the transducer with front-end electronic circuitry. The image reconstruction algorithm is based on phased subarray processing which allows for a low-cost front-end architecture while providing high-quality real-time images. The combination of CMUT technology and phased subarray (PSA) imaging will enable us to realize a small ultrasound scanner where the size and cost of the front-end circuitry is critical. We have achieved the following preliminary results: (1) Identified several image-guided surgical applications for cancer treatment, (2) specified the probe parameters based on physician input, (3) fabricated 128-element 1D CMUT array and (4) used it to form a high-quality 2D image, (4) fabricated a prototype 16x16-element 2D CMUT array, (5) developed PSA theory for 2D and 3D imaging, (6) tested 2D PSA theory with simulations and experimental data, and (7) shown that the system specs allow for real-time 3D ultrasound imaging. We propose to take the next major developmental step and demonstrate real-time 3D imaging from a miniature probe by accomplishing the following specific aims: (1) form preliminary 3D images using the prototype array and existing acquisition system, (2) design and fabricate new 16x16 2D CMUT arrays, (3) implement real-time beamforming and visualization, (4) design and build integrated circuits for the front-end probe electronics, (5) perform final integration into a fully function real-time 3D ultrasound system, and (6) perform preliminary clinical evaluation.

Agency
National Institute of Health (NIH)
Institute
National Cancer Institute (NCI)
Type
Exploratory/Developmental Grants Phase II (R33)
Project #
5R33CA099059-02
Application #
6771147
Study Section
Special Emphasis Panel (ZCA1-SRRB-9 (J2))
Program Officer
Farahani, Keyvan
Project Start
2003-07-02
Project End
2006-06-30
Budget Start
2004-07-01
Budget End
2005-06-30
Support Year
2
Fiscal Year
2004
Total Cost
$458,663
Indirect Cost
Name
Stanford University
Department
Type
Organized Research Units
DUNS #
009214214
City
Stanford
State
CA
Country
United States
Zip Code
94305
Kothapalli, Sri-Rajasekhar; Ma, Te-Jen; Vaithilingam, Srikant et al. (2012) Deep tissue photoacoustic imaging using a miniaturized 2-D capacitive micromachined ultrasonic transducer array. IEEE Trans Biomed Eng 59:1199-204
Khuri-Yakub, Butrus T; Oralkan, Omer (2011) Capacitive micromachined ultrasonic transducers for medical imaging and therapy. J Micromech Microeng 21:54004-54014
Khuri-Yakub, B T; Oralkan, Omer; Nikoozadeh, Amin et al. (2010) Miniaturized ultrasound imaging probes enabled by CMUT arrays with integrated frontend electronic circuits. Conf Proc IEEE Eng Med Biol Soc 2010:5987-90
Wygant, Ira O; Jamal, Nafis S; Lee, Hyunjoo J et al. (2009) An integrated circuit with transmit beamforming flip-chip bonded to a 2-D CMUT array for 3-D ultrasound imaging. IEEE Trans Ultrason Ferroelectr Freq Control 56:2145-56
Zhuang, Xuefeng; Wygant, Ira O; Lin, Der-Song et al. (2009) Wafer-bonded 2-D CMUT arrays incorporating through-wafer trench-isolated interconnects with a supporting frame. IEEE Trans Ultrason Ferroelectr Freq Control 56:182-92
Wygant, Ira O; Zhuang, Xuefeng; Yeh, David T et al. (2008) Integration of 2D CMUT arrays with front-end electronics for volumetric ultrasound imaging. IEEE Trans Ultrason Ferroelectr Freq Control 55:327-42
Zhuang, Xuefeng; Ergun, Arif S; Huang, Yongli et al. (2007) Integration of Trench-Isolated Through-Wafer Interconnects with 2D Capacitive Micromachined Ultrasonic Transducer Arrays. Sens Actuators A Phys 138:221-229
Bayram, Baris; Kupnik, Mario; Yaralioglu, Goksen G et al. (2007) Finite element modeling and experimental characterization of crosstalk in 1-D CMUT arrays. IEEE Trans Ultrason Ferroelectr Freq Control 54:418-30
Triadafilopoulos, George; Watts, H David; Higgins, Jack et al. (2007) A novel retrograde-viewing auxiliary imaging device (Third Eye Retroscope) improves the detection of simulated polyps in anatomic models of the colon. Gastrointest Endosc 65:139-44
Oralkan, Omer; Bayram, Baris; Yaralioglu, Goksen G et al. (2006) Experimental characterization of collapse-mode CMUT operation. IEEE Trans Ultrason Ferroelectr Freq Control 53:1513-23