This project needs to be reopened for the Microtron as soon as that machine is recommissioned. Due to the high single fraction doses employed in intraoperative radiotherapy, it is important to minimize areas of high dose. These areas arise due to the scattering relationship between the photon collimators and the applicator system. They may be non- symmetric due to either an asymmetric applicator design, or to manufacturing tolerances of the applicator system resulting in the center of the applicator to be slightly different than the central ray of the accelerator. These dose distributions can be optimized to predetermined criteria by the use of an optimun field size for each combination of applicator and electron energy. The use of asymmetric collimators can further correct for nonuniformity in the dose distribution due to asymmetric design, applicator bevel angle, or applicator mis-alignment.

Agency
National Institute of Health (NIH)
Institute
Division of Cancer Treatment (NCI)
Type
Intramural Research (Z01)
Project #
1Z01CM006377-04
Application #
3916592
Study Section
Project Start
Project End
Budget Start
Budget End
Support Year
4
Fiscal Year
1988
Total Cost
Indirect Cost
Name
Cancer Treatment
Department
Type
DUNS #
City
State
Country
United States
Zip Code