The overall goal of this project is to develop and commercialize a simple cell counter for routine enumeration of CD4 cells for management of HIV-infected individuals. HIV-1 mediated CD4 cell destruction is the central immunologic feature of HIV-1 infection. Thus, the CD4 count is a critical measurement in initial disease staging, in monitoring antiretroviral therapy and in managing primary and secondary prophylaxis for opportunistic infections. Flow cytometry is the current standard-of-care for CD4 cell counting. Unfortunately, flow is a central lab-based technique, and sample cold chain and transport, equipment, and operational costs render the technique cost-prohibitive in limited resource settings where HIV prevalence is highest. A simple, point-of-care (POC) CD4 counting tool will fundamentally change HIV-1 management. The first phases of this project have yielded a low-cost prototype instrument that delivers fast, accurate, and reliable CD4 counts for whole blood specimens from both HIV-infected individuals and healthy donors. Building on this success, we propose to complete the technical development by minimizing required user interactions with the device, driving toward an FDA CLIA-waived system.
The specific aims are: 1. Simplify the sample processing protocol and design and build a cartridge/instrument prototype for which the only user interaction is whole blood addition. We will investigate and develop assay reagent combinations (e.g., stain, lysis buffer, etc.), reagent packaging, and modified disposable cartridge and instrument designs to reduce the number of liquid transfer steps in the assay. The development goal is a CLIA-waived (or equivalent) system. 2. Incorporate in-cartridge quality controls that confirm correct function and user operation. We will develop a combination of calibration kit components and on-cartridge quality control features, including count calibrators, fluorescent stain indicators and image quality software flags. 3. Demonstrate the accuracy and usability of the CD4 counting system in clinical settings. Field evaluations of the prototype system will be performed at HIV-specific clinics affiliated with the University of California, San Diego (UCSD) Medical Center and in Maputo, Mozambique. Supporting this effort we have established strong clinical collaborations with global health thought leaders. Co-investigators Dr. Robert Schooley and Dr. Constance Benson at UCSD provide critical input on device requirements, and facilitate laboratory evaluations. At the conclusion of this project, the utility, value, and unique capabilities of the MBio point of care C4 cell counter will be strongly established, and the final steps in commercialization, i.e., productin instruments, clinical trials, regulatory submission, and marketing will be accelerated.

Public Health Relevance

HIV/AIDS remains a critical public health concern worldwide. It is estimated that by 2014, 12 million people will need access to affordable, point-of-care CD4 cell count measurements to effectively manage their HIV infection. The proposed research is aimed directly at providing this key medical technology.

Agency
National Institute of Health (NIH)
Institute
National Institute of Allergy and Infectious Diseases (NIAID)
Type
Small Business Innovation Research Grants (SBIR) - Phase II (R44)
Project #
5R44AI070052-05
Application #
8435377
Study Section
HIV/AIDS Vaccines Study Section (VACC)
Program Officer
Livnat, Daniella
Project Start
2006-03-01
Project End
2014-02-28
Budget Start
2013-03-01
Budget End
2014-02-28
Support Year
5
Fiscal Year
2013
Total Cost
$789,369
Indirect Cost
Name
Mbio Diagnostics, Inc.
Department
Type
DUNS #
961776577
City
Boulder
State
CO
Country
United States
Zip Code
80303
Givens, Monique; Weaver, Amanda; Bickman, Sarah et al. (2017) Near patient CD4 count in a hospitalized HIV patient population. Cytometry B Clin Cytom 92:451-455
Logan, Cathy; Givens, Monique; Ives, Jeffrey T et al. (2013) Performance evaluation of the MBio Diagnostics point-of-care CD4 counter. J Immunol Methods 387:107-13